Product Description
Note: this chamber comes with a FRESH CALIBRATION
Manufacturers Information:
Temperature Range – 73°C to +175°C
Control Tolerance ±0.5°C (±0.2°C Typical) (Measured at the control sensor after stabilization)
Uniformity ±1.0°C (Variations throughout the chamber after stabilization)
Input Power Requirements
230 V ±10%, 60 Hz, 3 PH Max Current Draw 39 A; Recommended Service 50 A
208 V -5/+10%, 60 Hz, 3 PH Max Current Draw 35 A; Recommended Service 45 A
Input may be configured for nominal 208 V or 230 V in the field by changing jumpers.
Call for other voltages or 50 Hz operation.
Workspace Dimensions 40" W x 32" H x 36.5" D (27 cubic feet)
Outside Dimensions 49" W x 73.25" H x 63" D (nominal)
Door latch adds 3" to width on right side. Circulator motor housing adds 6" to height.
Min. Installed Clearance 12" from the left and right side, 24" from the rear
Access Ports 4" Port on left and right side (two total), Supplied with foam plugs
Weight 1600 pounds
Manufacturers Information:
Temperature Range – 73°C to +175°C
Control Tolerance ±0.5°C (±0.2°C Typical) (Measured at the control sensor after stabilization)
Uniformity ±1.0°C (Variations throughout the chamber after stabilization)
Input Power Requirements
230 V ±10%, 60 Hz, 3 PH Max Current Draw 39 A; Recommended Service 50 A
208 V -5/+10%, 60 Hz, 3 PH Max Current Draw 35 A; Recommended Service 45 A
Input may be configured for nominal 208 V or 230 V in the field by changing jumpers.
Call for other voltages or 50 Hz operation.
Workspace Dimensions 40" W x 32" H x 36.5" D (27 cubic feet)
Outside Dimensions 49" W x 73.25" H x 63" D (nominal)
Door latch adds 3" to width on right side. Circulator motor housing adds 6" to height.
Min. Installed Clearance 12" from the left and right side, 24" from the rear
Access Ports 4" Port on left and right side (two total), Supplied with foam plugs
Weight 1600 pounds