Product Description
Accessories included:
Charge plate extension cable (5ft.)
Ground cable
Power cord
Operator’s Manual and case
SPECIFICATIONS
All specifications are referred to PLate VColtage unless otherwise specified.
Charging Voltage: ±10 to ±1000 volts, differential
adjustable 10 volts to 100 volts in 1 volt increments
Zero Stability: <100mV/sec
Timer: 0.1 to 999.9 seconds in 0.1 sec increments
1000 to 9999 seconds in 1 sec increments
Start Voltages: 1000 volts ±0.3% standard, adjustable
between 10 and 1000 volts in 1 volt increments
Stop Voltages: 100 volts ±3% standard, adjustable
between 0 and 995 volts in 1 volt increments
Peak Displays: ±peak voltage during float mode
Power: 90-250 VAC 50/60 Hz, internal 12 volt
battery with built-in charger (6hr/charge)
Voltage Output Monitor:divide by 200, 0.1%, speed of response
<10ms
Display: 240 x 64, character/graphic Voltage Display:
3 ½ digit, ±1.0 volt resolution
Timer Display: 4 digit
Accuracy: electrometer ±0.1% of reading, ±1 volt
referred to input Bandwidth: 1kHz at 20V p-p , 10Hz at 2000V p-p
Zero Drift: <100mV/sec
Charge Plate Capacitance: 20pF, ±5%
Plate Self-discharge: <200mV/sec
Operation Temperature: 5° C to 35°C
Temperature Sensor: ±2° C typ
Operating RH: to 80%, non-condensing
Humidity Sensor: ±5% typ from 10% to 80% RH @25° C