Part Number: sku7805
US$275.00
Add Items to Cart
Quantity in Stock: 11

Ask a Question

Product Description

Inficon PCG410 / 355-020 / 355020 Pirani Capacitance Diaphragm Gauge

The PCG gauge is a combination gauge consisting of a Pirani sensor and a capacitive diaphragm sensor. Both
sensors are constantly active. At low pressures, only the signal of the Pirani sensor is used for pressure measurement; at high pressures, only the signal of the capacitive diaphragm sensor. To determine the output
signal in the intermediate range, both signals are used proportional to the pressure.

Technical Data:

Measurement principle
100 mbar 1) … 1500 mbar                  capacitance diaphragm sensor
5×10-4 … 1 mbar                                thermal conductance according to Pirani
1 … 100 1) mbar                                 crossover range

Measurement range                           5×10-4 … 1500 mbar
Accuracy (N2)
1×10-3 … 50 mbar                            ±15% of reading
50 … 950 mbar                                 ±5% of reading         

Atmospheric pressure                       ±2.5% of reading
(950 … 1050 mbar)

Repeatability (N2)                             ±2% of reading
                                                          (1×10-3 … 1100 mbar)

Output signal (measurement signal)
Voltage range                                     0 … +9.0 V
Mesasurement range                        +2.2 … +8.68 V
Voltage vs. pressure                          1 V/decade, logarithmic
Output impedance                              2 × 4.7 Ω, short circuit-proof
Minimum load impedance                  10 kΩ
Response time                                   10 ms
HV Adjustment at                               <<10-4 mbar(with potentiometer <HV>)

Internal Volume: 6cm3
Weight: 90g

*Full Spec Sheet available upon request